The DTFB Limited Access Digital Tension meter accurately measure the running line tensions of yarns, fibers, wires, cables and similar process materials. Long, flat fingers and miniature 7.4 mm diameter V-grooved rollers permit measurement in areas with limited space or tightly grouped filaments.
DTFX Additional Features
- Built-in datalogging with RS232 and 0-1 VDC Analog output
- Memory capacity of 100 values plus min, max, peak, avg and standard deviation
- Optional TensionView software available for batch download or continuous real-time trending
- Roller shifting mechanism ensures easy acquisition of the running material
- Small measuring head, allows access where space is limited or where filaments run close together
- Turned-up outer finger edges guide the running filament into the roller grooves
- Calibration to customer supplied material is available
- Minature 7.4 mm diameter V-Grooved rollers for material speeds up to 900 meters/min
Model | Range | Resolution | Measuring Head Width mm |
---|---|---|---|
DTFB-200 DTFX-200 | 2.0-200.0 g | 0.1 g | 38 |
DTFB-500 DTFX-500 | 5.0-500.0 g | 0.1 g | 38 |
DTFB-1K DTFX-1K | 50-1000 g | 1 g | 36 |
DTFB-2K DTFX-2K | 200-2000 g | 1 g | 36 |
Other units of measure available.
Order Code | Line Speed ...m/min | Roller Material |
---|---|---|
Standard | 900 | Hardcoated aluminium |
Code K | 2000 | Hardcoated aluminum |
Code T | 450 | Plastic (POM) black |
Code W | 450 | Nickel-plated steel |
Accuracy | From 10% up to 90% of tension range: ±0.5% full scale and ±1 digit remaining tension range and other calibration material: ± 3% full scale and ±1 digit |
Overrange | Approx. 15% full scale, without accuracy guarantee |
Overload protection | 100% of tension range |
Measuring principle | Strain gauge bridge |
Measuring roller deflection | 0.2 mm max. |
Converter | 12 bit A/D |
Temperature Coefficient | Gain: less than ±0.01 % full scale/ºC Zero point: less than ±0.3 % full scale/ºC |
Display | LCD 4 digit, 12 mm high |
Display update rate | Adjustable 0.5, 1, 2 or 4 s (averaging) |
Field adjustment | ±7 steps in 1.5 % increment |
Memory | MIN, MAX, PEAK and last value |
Temperature range | 10 -45 ºC |
Air humidity | 85 % relative humidity, max. |
Housing Dimensions | 235 x 76 x 45mm (LxWxH) |
Additional specifications model for DTFX | |
---|---|
Expanded memory | Max. 100 readings, MAX, MIN, and PEAK as well as average values and standard deviation |
Digital output signal | RS 232 C (4800, 8, N, 2) |
Analog output signal | 0 - 1 V DC (converter rate 16 ms) |
Digimatic interface | Mitutoyo |
Thank You
Thank you for your quote request, we will contact you within 1-2 business days.
There was an error submiting the form. Please try again.