0.0060" - 1.000" / 0.150 - 25.40 mm
Range with Contact Transducers
up to 36.00" / 915.0 mm
The PZX-7 is a precision ultrasonic wall thickness gauge designed for high-resolution measurements on thin-wall metal and plastics. It is packaged in a sturdy extruded aluminum housing which is sealed for excellent environmental protection assuring trouble-free use even in the toughest field and production environments.
The PZX-7 has measuring range in steel from 0.0060 to 1.0000 inch, (0.15 to 25.40 mm). Using a single element delay tip transducer, the gauges will measure thin materials and automatically remove any coatings in Echo-Echo Mode (E-E) as well as automatically switch to Interface Echo Mode (I-E)when measuring thicker materials, For measurement of plastics, the user would select Plastic Mode (PLAS) and use the option Graphite Delay Line Tip (p/n X-633-0000).
Additional Measuring Modes include Differential (DIFF) where a user sets a "nominal value" and the gauge displays the ± difference between then nominal and actual wall thickness, Alarm (user-set Hi/Low Limits) trigger Hi/Lo LED's when the measured values are out of tolerance and Velocity Mode
The PZX-7 can also be used with single element contact transducers for wall thickness measurements up to 36 inches (923 mm). Although the PZX-7 does not have any built-in memory, it can transfer single values (value on the display) via the USB port to a PC or compatible data collector. No Software is required.
- Resolution of 0.0001 inch (0.001 mm)
- Large, backlit LCD display retains last reading
- Selectable units of inch or mm
- Selectable Backlight ON/OFF/AUTO
- Includes NIST Traceable Calibration Certificate
- IP 65 Rated for protection even in the toughest field conditions
- 5 Year Warranty, CE-Certified and Made in USA
- Scan mode, 100 readings/sec. Probe is slid over a large measuring area, the minimum thickness during the "scan" will be displayed
- Adjustable GAIN (AGC) in 5 steps to fine tune the gauge for use on challenging materials or applications
- Can be used with both Single Element Delay Line and contact transducers (5 to 20MHz)
- LCD shows Thickness Value, Velocity Setting, Gain Setting, Stability & Battery Indicators, Scan Mode, Zero and units
- Operates on two (2) AA Batteries - 45 hours continuous operation
- Calibrate to a sample of known thickness which automatically calculates the optimal acoustic velocity
- USB Data output of single values for transfer to data collection devices (no internal memory)
- Powered by: 120MHz FPGA timing & 100 volt spike pulser
- Measure Modes: Pulse-Echo (P-E), Echo-Echo (E-E), Interface Echo (I-E) and Plastics (PLAS)
- Optional serial RS-232 or Bluetooth output
- Additional Modes Include Differential (from user-entered "nominal value"), Alarm (user-set Hi/Low Limits) and Velocity Mode
P/N | Description | Standard / Optional | Image |
F-000-7102 | 1/4" x 3/8" for all metals (Short) |
Standard | |
F-000-7102 | 1/4" x 1/2" for all metals (Long) |
Standard | |
X-633-0000 | 1/4" x 3/8" Graphite Tip for thin plastics (Graphite Short) |
OPTIONAL |
Range with Standard Probe | 0.0060-1.0000" (0.15-25.40 mm) with standard delay line transducer in steel |
Resolution | 0.0001" (0.001 mm) . 0.001" (0.01 mm), user-selected units and resolution |
Display | Multi-function 7 segment 4.5 digit liquid crystal display with Back-light |
Velocity Range | 0.0120 to .7300 in/μs (305 to 18,542 m/sec) |
Display Backlight | Backlight is selectable on/off/auto, and selectable brightness (Lo, Med, Hi) options |
Output Optional | RS-232 and Bluetooth |
Cable | 4 ft. (1.2 m) with Microdot/Lemo00 connector. |
Output | USB single value only (no internal memory)TI-007DLX for a model with build-in memory for 40,000 values USB output |
Operating Temperature | -22 to 167F (-30 to 75C) |
Housing Protection | IP65 |
Battery Life | Typically operates for 35 hours on alkaline and 18 hours on NiCad |
Weight | 11 ounces (308 grams) |
Dimensions | 2.5 x 5.17 x 1.25" (64 x 131 x 32 mm) |
Warranty | Gauge: 5 Years Probes: 90 Days |
Total Measuring Range (Steel) | 0.006 - 36.000 inches 0.15 - 924 mm (depends on material and transducer/probe type) |
Measuring Modes | Pulse-Echo (P-E), Echo-Echo (E-E), Interface Echo (I-E) and Plastics (PLAS), Differential, Alarm, Scan and Velocity (VX) |
Gain | Automatic time dependent gain (TDG) with manual override |
Probe (Standard) | 1/4" diameter, 15 MHz delay line transducer with cable (p/n T-402-5507) |
Probes (optional) | Various Single Element Delay Line and Contact Transducers (5 to 20MHz) |
Display Update Rate | 10 Hz (10 updates/sec) |
Battery Type | 2x AA 1.5V alkaline, 1.2V NiCad, or 1.5V lithium AA cells (rechargeable batteries can be used) |
Housing | Extruded aluminum body with nickel-plated aluminum end caps (gasket sealed) |
Certifications | NIST Traceable and MIL-STD-45662A |
Keypad | Sealed membrane that is resistant to both water and petroleum products Seven or eight tactile-feedback keys |
Pulse Repetition Frequency (PRF) | 200 Hz (200 pulses/sec) |
To Lock Settings | With gauge powered off, press and hold the "+" key while powering on the gauge The LOCK icon will be illuminated on the Display |
To unLock Settings | With gauge powered off, press and hold the "-" key while powering on the gauge The LOCK icon will turn off |
Item | Description | Price | ||
---|---|---|---|---|
Dakota PZX-7 / PZX1 Precision Ultrasonic Wall Thickness Gauge kit with T-402-5507 probe, Z-307-0001, Z-307-0002 | $2,250.00 | |||
Accessories | ||||
Protective Rubber Boot for TI / ZX Series Small Body Ultrasonic Thickness Gauges | $95.00 | |||
Delay Line Tips | ||||
F-000-7102 | 1/4" x 3/8" Length Delay Line - OLD PN X-231-0000 | $20.00 | ||
F-000-7103 | 1/4" Diameter x 1/2" Length Delay Line Tip - OLD PN X-241-0000 | $20.00 | ||
X-633-0000 | 1/4" Diameter x 3/8" Length Graphite Tip Delay Line | $50.00 | ||
Spare / Replacement Delay Line Transducer | ||||
1/4" diameter, 15 MHz delay line transducer with cable, 1/2" and 3/8" length delay line tips | $615.00 | |||
Single Element Contact Transducers - for Thick Walls (supplied without cable) | ||||
T-4023-2855 | 0.250", 5 Mhz, Side-in, Microdot Non-Magnetic Single Element Contact Transducer | $375.00 | ||
T-4043-2855 | 0.5", 5 Mhz, Side-in, Microdot Non-Magnetic Single Element Contact Transducer | $375.00 | ||
Other Accessories for PZX-7 | ||||
LEMO 00 - Microdot Single Probe Transducer Cable, 4 foot length | $80.00 | |||
BATT-RC4AA-WC | 4-pack AA Ni-MH Rechargeable Batteries & Charger Kit | $35.00 | ||
A-303-1002 | Bluetooth Option for ZX-5, ZX-6, PZX-7 (TI-007X) - for connection to a PC | $225.00 | ||
A-303-1003 | Bluetooth Option for ZX-5, ZX-6, PZX-7 (TI-007X) - for connection to the APP | $225.00 |