6000GP Series General Purpose Models |
Range: 0 - 60 mils (1500 µm)
Resolution: 0.05 mil / 0.1 mil (1 µm / 2 µm)
Models: 6000-FS1, 6000-FS3, 6000-NS1, 6000-NS3, 6000-FNS1, 6000-FNS3, 6000-FRS1, 6000-FRS3, 6000-NRS1, 6000-NRS3, 6000-FNRS1, 6000-FNRS3
The 6000GP Series General Purpose Coating Thickness Gauge is a precise, rugged tester designed to measure coating thickness on ferrous and or non-ferrous metals accurately and quickly using magnetic and eddy current principles
The 6000GP is available in Standard (1) or Advanced (3) configuration. The standard model has a monochrome display, 250 reading memory and USB data output. The Advanced model adds a high contrast color LCD display, Scan Mode, Real-Time Graphing Bluetooth output and more. |

Easy-To-Use
- Ready to measure—no calibration adjustment required for most applications
- Enhanced one-handed menu navigation
- Flashing display—ideal in a noisy environment
- RESET feature instantly restores factory settings
Durable
- Solvent, acid, oil, water and dust resistant - weatherproof
- Wear-resistant probe tip
- Shock-absorbing, protective rubber holster with belt clip
- Two year warranty on gage body AND probe
Accurate
- Certificate of Calibration showing traceability to NIST included (Long Form)
- Built-in temperature compensation ensures measurement accuracy
- Hi-RES mode increases displayed resolution for use on applications that require greater precision
- Conforms to national and international standards including ISO and ASTM
Versatile
- Fully interchangeable probes are compatible with ALL gage bodies
- Multiple Calibration adjustment options including 1 point, 2 point, known thickness, average zero
- Selectable display languages
- Hi Contrast backlit display for bright or dark environments
- Flip display enables right-side-up viewing
- Extended cables available (up to 75 m/250 ft) for underwater or remote measuring
- Uses alkaline or rechargeable batteries (built-in charger)
Powerful
- Statistics mode continually displays/updates average, standard deviation, min/max thickness and number of readings while measuring
- HiLo alarm audibly and visibly alerts when measurements exceed user-specified limits
- FAST mode - faster measurement speed for quick inspection
- USB port for fast, simple connection to a PC and to supply continuous power. USB cable included
- USB mass storage - stored readings and graphs can be accessed using universal PC/ Mac web browsers or file explorers. No software required
- Every stored measurement is date and time stamped
- Software updates via web keep your gage current
- Connects to PosiTector.net
|
|
|
Probe Models: |
FS, NS, FNS |
Range: |
0 -60 mils |
| 0 - 1500 um |
Accuracy*: |
+ (0.05 mils + 1%) 0 - 2 mils
+ (0.1 mil + 1%) > 2 mils |
+ (1 um + 1%) 0 - 50 um
+ (2 um + 1%) > 50 um |
| *Accuracies are stated as a fixed value plus a percentage of the gage's actual reading. |
|
|
Probe Models: |
FRS, NRS, FNRS |
Range: |
0 -60 mils |
| 0 - 1500 um |
Accuracy*: |
+ (0.05 mils + 1%) 0 - 2 mils
+ (0.1 mil + 1%) > 2 mils |
+ (1 um + 1%) 0 - 50 um
+ (2 um + 1%) > 50 um |
| *Accuracies are stated as a fixed value plus a percentage of the gage's actual reading. |


|
Feature Comparison Chart
|
Basic
(1)
|
Memory (3)
|
| Hi Contrast LCD Display |

Monochrome
|

Color
|
| FAST mode - faster measurement speed for quick inspection |
|
|
| SCAN Mode - take continuous readings without lifting the probe |
-
|
|
| Hi-Lo limits audibly and visibly alerts when
measurements exceed your specified limits |
|
|
| Instantly calculate average, standard
deviation, min/max thickness and number of readings while measuring |
|
|
| Reading Storage Capacity |
250 in one group
|
10,000 in up to 1000 groups
|
| SSPC PA2 feature set determines if film
thickness over a large area conforms to user-specified min/max levels |
-
|
|
| PSPC 90/10 feature determines if a coating
system complies with an IMO performance standard for protective coatings |
-
|
|
| Onscreen help, real time graphing, picture prompting and batch notes |
-
|
|
| Multiple stored calibration adjustments for
measuring on a variety of substrate conditions |
-
|
|
| Wireless Technology for data transfer to a PC or optional portable printer |
-
|
|
|
|
Measuring principle:
|
magnetic induction principle (Ferrous) eddy-current principle (Non-Ferrous)
|
|
Display
|
Backlit & Reversible LCD
|
|
Data Output
|
USB - "1" & "3" Models Bluetooth - "3" Models
|
|
Surface Temp.
|
Maximum 120 °F / 49 °C
|
|
Power
|
3 AAA batteries, Included
|
|
Dimensions
|
5.75" x 2.5" x 1.2" (146 x 64 x 31 mm)
|
|
Weight
|
5.8 oz. (165 g) without batteries
|
|
Protection Class
|
Meets or exceeds IP5X, Solvent, acid, oil, water & dust resistant
|
|
Warranty
|
2 Years
|
|
Calibration Certificate
|
Certificate of Calibration showing traceability to NIST included
|
|
Min. Concave Radius
|
1 inch (2" OD)
|
|
6000IP Series Intergral Probe Models |  |
Integral probe for one hand use on paint and coating applications in the range of 0-60 mils (0-1500 µm)
| | Click image for more info | | |
|
6000MP Series Micro Probe Models |  |
Micro Probe for measuring thin coatings, small parts or hard to reach areas or spots
| | Click image for more info | | |
|
6000TC Series Thick Coating Model |  |
For thick paint and coatings from 60 to 250 mils
| | Click image for more info | | |
|
6000TCX Series Extra Thick Coating Models |  |
For extra thick paint and coatings from 150-500 mils
| | Click image for more info | | |
|
PRINTERBT Bluetooth Printer |  |
Battery-operated
Receives data via Bluetooth wireless technology
Prints readings and statistical summary
| | Click image for more info | | |
|
PosiTectorNET
|  |
A free web-based application offering secure centralized management of thickness readings
| | Click image for more info | | |
|
PRBFOOTKIT-D Probe Foot Rings |  |
Probe foot rings, ideal for use on thin, soft coatings.
Protects coatings from unnecessary wear while using Scan or Fast mode to measure large objects.
| | Click image for more info
|
| | | | ALL Gauges are supplied as a complete kit including precision plastic shims, protective rubber holster with belt clip, wrist strap, 3 AAA alkaline batteries, instructions, nylon carrying case with shoulder strap, protective lens shield, Long Form Certificate of Calibration traceable to NIST, USB cable, two (2) year warranty. |
 |
|
|
|